laboneq_applications.contrib.experiments.two_qubit.chevron
¶
This module defines a CZ chevron experiment on a pair of qubits.
Each pair, written here as q0 and q1, is coupled through a tunable coupler (TC).
In this experiment, we sweep the duration of the fast flux pulses applied to the
second qubit of the pair and to the TC. The amplitude of one of those two flux
pulses is swept as well: the pulse that drives the gate, which is the qubit
pulse if the edge defines a control_pulse and the TC pulse if it does not.
The chevron experiment has the following pulse sequence:
q0 -[pi]---------------[measure][reset]---
q1 -[pi][modulated ffp][measure][reset]---
tc -----[ ffp ]---------[discharge]----------
create_experiment(qpu, qubit_pairs, durations, amplitudes, options=None)
¶
Creates an Cz tuneup experiment workflow.
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
qpu
|
QPU
|
The qpu consisting of the original quibts, couplers, and quantum operations. |
required |
qubit_pairs
|
list[list[str]]
|
Pairs of qubits to run the experiment on. |
required |
durations
|
QubitSweepPoints
|
The durations of the fast flux pulses on the couplers or control qubit. Must be a list of numbers or an array. |
required |
amplitudes
|
QubitSweepPoints
|
The amplitues of the fast flux pulses on the coupler or control qubit. Must be a list of numbers or an array. |
required |
options
|
TuneupExperimentOptions | None
|
The options for building the experiment. See TuneupExperimentOptions for accepted options. |
None
|
Returns:
| Name | Type | Description |
|---|---|---|
experiment |
Experiment
|
The generated LabOne Q experiment instance to be compiled and executed. |
experiment_workflow(session, qpu, qubit_pairs, durations, amplitudes, temporary_parameters=None, options=None)
¶
The Chevron Workflow.
The workflow consists of the following steps:
Parameters:
| Name | Type | Description | Default |
|---|---|---|---|
session
|
Session
|
The connected session to use for running the experiment. |
required |
qpu
|
QPU
|
The qpu consisting of the original qubits, couplers and quantum operations. |
required |
qubit_pairs
|
ArrayLike[ArrayLike[QuantumElement, QuantumElement]]
|
Pairs of qubit to run the experiment on. |
required |
durations
|
QubitSweepPoints
|
The durations of the fast flux pulses on the couplers or control qubit. Must be a list of numbers or an array. |
required |
amplitudes
|
QubitSweepPoints
|
The amplitues of the fast flux pulses on the coupler or control qubit. Must be a list of numbers or an array. |
required |
temporary_parameters
|
dict[str, dict | TunableTransmonQubitParameters] | None
|
The temporary parameters to update the qubits and couplers with. |
None
|
options
|
TuneUpWorkflowOptions | None
|
The options for building the workflow. In addition to options from WorkflowOptions |
None
|
Returns:
| Name | Type | Description |
|---|---|---|
result |
None
|
The result of the workflow. |