Troubleshooting

This section aims to help the user solve and avoid problems while using the software and operating the instrument.

Common Problems

Your SHFQC Instrument is an advanced piece of laboratory equipment which has many more features and capabilities than a traditional signal generator. In order to benefit from these, the user needs access to a large number of settings in the API or the LabOne User Interface. The complexity of the settings might overwhelm a first-time user, and even expert users can get surprised by certain combinations of settings. This section provides an easy-to-follow checklist to solve the most common mishaps.

Table 1: Common Problems
Problem Check item
The software cannot be installed or uninstalled Please verify you have administrator/root rights.
The software cannot be updated Please use the Modify option in Windows Apps & Features functionality. In the software installer select Repair, then uninstall the old software version, and install the new version.
The Instrument does not turn on Please verify the power supply connection and inspect the fuse. The fuse holder is integrated in the power connector on the back panel of the instrument.
The Instrument can’t be connected over USB Please verify that the USB port labeled "Maintenance" is connected. The port labeled "USB" is not currently supported and will be enabled with a future LabOne release.
The Instrument performs close to specification, but higher performance is expected After 2 years since the last calibration, a few analog parameters are subject to drift. This may cause inaccurate measurements. Zurich Instruments recommends re-calibration of the Instrument every 2 years.
The Instrument measurements are unpredictable Please check the Status Tab to see if there is any active warning (red flag), or if one has occurred in the past (yellow flag).
The Instrument does not generate any output signal Verify that the signal output switch of the right signal output channel has been activated in the Output tab.
The LabOne User Interface does not start Verify that the LabOne Data Server (ziDataServer.exe) and the LabOne Web Server (ziWebServer.exe) are running via the Windows Task Manager. The Data Server should be started automatically by ziService.exe and the Web Server should be started upon clicking "Zurich Instruments LabOne" in the Windows Start Menu.
If both are running, but clicking the Start Menu does not open a new User Interface session in a new tab of your default browser then try to create a new session manually by entering 127.0.0.1:8006 in the address bar of your browser.
The user interface does not start or starts but remains idle Verify that the Data Server has been started and is running on your host computer.
The user interface is slow and the web browser process consumes a lot of CPU power Make sure that the hardware acceleration is enabled for the web browser that is used for LabOne. For the Windows operating system, the hardware acceleration can be enabled in Control Panel → Display → Screen Resolution. Go to Advanced Settings and then Trouble Shoot. In case you use a NVIDIA graphics card, you have to use the NVIDIA control panel. Go to Manage 3D Settings, then Program Settings and select the program that you want to customize.

Location of the Log Files

The most recent log files of the LabOne Web and Data Server programs are most easily accessed by clicking on in the LabOne Device Connection dialog of the user interface. The Device Connection dialog opens on software start-up or upon clicking on in the Config tab of the user interface.

The location of the Web and Data Server log files on disk are given in the sections below.

Windows

The Web and Data Server log files on Windows can be found in the following directories.

  • LabOne Data Server (ziDataServer.exe):

    C:\Windows\ServiceProfiles\LocalService\AppData\Local\Temp\Zurich Instruments\LabOne\ziDataServerLog

  • LabOne Web Server (ziWebServer.exe):

    C:\Users[USER]\AppData\Local\Temp\Zurich Instruments\LabOne\ziWebServerLog

Note

The C:\Users\[USER]\AppData folder is hidden by default under Windows. A quick way of accessing it is to enter %AppData%\.. in the address bar of the Windows File Explorer.

Figure 1: Using the

Linux and macOS

The Web and Data Server log files on Linux or macOS can be found in the following directories.

  • LabOne Data Server (ziDataServer):

    /tmp/ziDataServerLog_[USER]

  • LabOne Web Server (ziWebServer):

    /tmp/ziWebServerLog_[USER]

Prevent web browsers from sleep mode

It often occurs that an experiment requires a long-time signal acquisition; therefore, the setup including the measurement instrument and LabOne software are left unattended. By default, many web browsers go to a sleep mode after a certain idle time which results in the loss of acquired data when using the web-based user interface of LabOne for measurement. Although it is recommended to take advantage of LabOne APIs in these situations to automate the measurement process and avoid using web browsers for data recording, it is still possible to adjust the browser settings to prevent it from entering the sleep mode. Below, you will find how to modify the settings of your preferred browser to ensure a long-run data acquisition can be implemented properly.

Edge

  1. Open Settings by typing edge://settings in the address bar

  2. Select System from the icon bar.

  3. Find the Never put these sites to sleep section of the Optimized Performance tab.

  4. Add the IP address and the port of LabOne Webserver, e.g., 127.0.0.1:8006 or 192.168.73.98:80 to the list.

Chrome

  1. While LabOne is running, open a tab in Chrome and type chrome://discards in the address bar.

  2. In the shown table listing all the open tabs, find LabOne and disable its Auto Discardable feature.

  3. This option avoids discarding and refreshing the LabOne tab as long as it is open. To disable this feature permanently, you can use an extension from the Chrome Webstore.

Firefox

  1. Open Advanced Preferences by typing about:config in the address bar.

  2. Look for browser.tabs.unloadOnLowMemory in the search bar.

  3. Change it to false if it is true.

Opera

  1. Open Settings by typing opera://settings in the address bar.

  2. Locate the User Interface section in the Advanced view.

  3. Disable the Snooze inactive tabs to save memory option and restart Opera.

Safari

  1. Open Debug menu.

  2. Go to Miscellaneous Flags.

  3. Disable Hidden Page Timer Throttling.