Troubleshooting
This section aims to help the user solve and avoid problems while using the software and operating the instrument.
Common Problems
Your GHFLI Instrument is an advanced piece of laboratory equipment which has many more features and capabilities than a traditional lock-in amplifier. In order to benefit from these, the user needs access to a large number of settings in the LabOne User Interface. The complexity of the settings might overwhelm a first-time user, and even expert users can get surprised by certain combinations of settings. To avoid problems, it’s good to use the possibility to save and load settings in the Config Tab. This allows one to keep an overview by operating the instrument based on known configurations. This section provides an easy-to-follow checklist to solve the most common mishaps.
Problem | Check item |
---|---|
The software cannot be installed or uninstalled |
Please verify you have administrator/root rights. |
The software cannot be updated |
Please use the Modify option in Windows Apps & Features functionality. In the software installer select Repair, then uninstall the old software version, and install the new version. |
The Instrument does not turn on |
Please verify the power supply connection and inspect the fuse. The fuse holder is integrated in the power connector on the back panel of the instrument. |
The Instrument can’t be connected over USB |
Please verify that the instrument is connected through the "USB 1" port. The port labeled "USB 2" is not currently supported and will be enabled with a future LabOne release. |
The Instrument has a high input noise floor (when connected to host computer by USB) |
the USB cable connects the Instrument ground to computer ground, which might inject some unwanted noise to the measurements results. In this case it is recommended to use the Ethernet connection which is galvanically isolated using a UTP Cat 5 or 6 cable (UTP stands for "unshielded twisted pair"). |
The Instrument performs poorly at low frequencies (below 100 kHz) : |
the signal inputs of the instrument might be set to AC operation. Please verify to turn off the AC switch in the Lock-in Tab or In / Out Tab. |
The Instrument performs poorly during operation |
the demodulator filters might be set too wide (too much noise) or too narrow (slow response) for your application. Please verify if the demodulator filter settings match your frequency versus noise plan. |
The Instrument performs poorly during operation |
clipping of the input signal may be occurring. This is detectable by monitoring the red LEDs on the front panel of the instrument or the Input Overflow (OVI) flags on the STATUS_TAB of the user interface. It can be avoided by adding enough margin on the input range setting (for instance 50% to 70% of the maximum signal peak). |
The Instrument performs strangely when working with the GHF-MF Multi-frequency Option |
it is easily possible to turn on more signal generators than intended. Check the generated Signal Output with the integrated oscilloscope and check the number of simultaneously activated oscillator voltages. |
The Instrument performs close to specification, but higher performance is expected |
After 2 years since the last calibration, a few analog parameters are subject to drift. This may cause inaccurate measurements. Zurich Instruments recommends re-calibration of the Instrument every 2 years. |
The Instrument measurements are unpredictable |
Please check the Status Tab to see if there is any active warning (red flag), or if one has occurred in the past (yellow flag). |
The Instrument does not generate any output signal |
verify that signal output switch has been activated in the Lock-in Tab or in the In / Out Tab. |
The sample stream from the Instrument to the host computer is not continuous |
Check the communication (COM) flags in the status bar. The three flags indicate occasional sample loss, packet loss, or stall. Sample loss occurs when a sampling rate is set too high (the instrument sends more samples than the interface and the host computer can absorb). The packet loss indicates an important failure of the communications to the host computer and compromises the behavior of the instrument. Both problems are prevented by reducing the sample rate settings. The stall flag indicates that a setting was actively changed by the system to prevent UI crash. |
The LabOne User Interface does not start |
Verify that the LabOne Data Server
( If both are running, but clicking the Start Menu does not open a new User Interface session in a new tab of your default browser then try to create a new session manually by entering |
The user interface does not start or starts but remains idle |
Verify that the Data Server has been started and is running on your host computer. |
The user interface is slow and the web browser process consumes a lot of CPU power |
Make sure that the hardware acceleration is enabled for the web browser that is used for LabOne. For the Windows operating system, the hardware acceleration can be enabled in |
Location of the Log Files
The most recent log files of the LabOne Web and Data Server programs are most easily accessed by clicking on in the
of the user interface. The Device Connection dialog opens on software start-up or upon clicking on
in the Config tab of the user interface.
The location of the Web and Data Server log files on disk are given in the sections below.
Windows
The Web and Data Server log files on Windows can be found in the following directories.
-
LabOne Data Server (
ziDataServer.exe
):C:\Windows\ServiceProfiles\LocalService\AppData\Local\Temp\Zurich Instruments\LabOne\ziDataServerLog
-
LabOne Web Server (
ziWebServer.exe
):C:\Users\[USER]\AppData\Local\Temp\Zurich Instruments\LabOne\ziWebServerLog
The ![]() Figure 1. Using the
%AppData%\.. shortcut in Windows Explorer to access the hidden folder. |
Prevent web browsers from sleep mode
It often occurs that an experiment requires a long-time signal acquisition; therefore, the setup including the measurement instrument and LabOne software are left unattended. By default, many web browsers go to a sleep mode after a certain idle time which results in the loss of acquired data when using the web-based user interface of LabOne for measurement. Although it is recommended to take advantage of LabOne APIs in these situations to automate the measurement process and avoid using web browsers for data recording, it is still possible to adjust the browser settings to prevent it from entering the sleep mode. Below, you will find how to modify the settings of your preferred browser to ensure a long-run data acquisition can be implemented properly.
Edge
-
Open Settings by typing
edge://settings
in the address bar -
Select System from the icon bar.
-
Find the Never put these sites to sleep section of the Optimized Performance tab.
-
Add the IP address and the port of LabOne Webserver, e.g.,
127.0.0.1:8006
or192.168.73.98:80
to the list.
Chrome
-
While LabOne is running, open a tab in Chrome and type
chrome://discards
in the address bar. -
In the shown table listing all the open tabs, find LabOne and disable its Auto Discardable feature.
-
This option avoids discarding and refreshing the LabOne tab as long as it is open. To disable this feature permanently, you can use an extension from the Chrome Webstore.
Firefox
-
Open Advanced Preferences by typing
about:config
in the address bar. -
Look for
browser.tabs.unloadOnLowMemory
in the search bar. -
Change it to false if it is true.