Table 1: General and storage 
| Parameter | 
Min | 
Typ | 
Max | 
| storage temperature | 
–25 °C | 
- | 
65 °C | 
| storage relative humidity (non-condensing) | 
- | 
- | 
95% | 
| operating temperature | 
5 °C | 
- | 
40 °C | 
| operating relative humidity (non-condensing) | 
- | 
- | 
90% | 
| specification temperature | 
18 °C | 
- | 
28 °C | 
| power consumption | 
- | 
- | 
120 W | 
| operating environment | 
IEC61010, indoor location, installation category II, pollution degree 2 | 
  | 
  | 
| operating altitude | 
up to 2000 meters | 
  | 
  | 
| power inlet fuses | 
250 V, 2 A, fast acting, 5 x 20 mm | 
  | 
  | 
| power supply AC line | 
100-240 V (±10%), 50/60 Hz | 
  | 
  | 
| dimensions (width x depth x height) | 
44.0 × 38.0 × 4.4 cm (no handle), 17.3 × 15.0 × 1.7 inch, 19 inch rack compatible | 
  | 
  | 
| weight | 
6 kg (14 lb) | 
  | 
  | 
| recommended calibration interval | 
2 years | 
  | 
  | 
Table 2: Maximum ratings 
| Parameter | 
Details | 
Min | 
Typ | 
Max | 
| Damage threshold Pump Out | 
voltage level | 
-15 V | 
- | 
+15 V | 
| Damage threshold Pump In | 
voltage level | 
-15 V | 
- | 
+15 V | 
|   | 
RF power | 
- | 
- | 
+20 dBm | 
| Damage threshold Signal Out | 
voltage level | 
-15 V | 
- | 
+15 V | 
| Damage threshold Signal In | 
voltage level | 
-15 V | 
- | 
+15 V | 
|   | 
RF power | 
- | 
- | 
+20 dBm | 
| Damage threshold Trig Out | 
voltage level | 
-0.5 V | 
- | 
+3.8 V | 
| Damage threshold Trig In | 
voltage level (50 Ω input impedance) | 
-0.5 V | 
- | 
+5.5 V | 
| Damage threshold DIO Out | 
voltage level | 
-0.5 V | 
- | 
+3.8 V | 
| Damage threshold DIO In | 
voltage level | 
-0.5 V | 
- | 
+6.5 V | 
| Damage threshold External Clk Out | 
voltage level | 
-3 V | 
- | 
+3 V | 
| Damage threshold External Clk In | 
voltage level | 
-15 V | 
- | 
+15 V | 
|   | 
RF power | 
- | 
- | 
+15 dBm | 
Table 3: Host computer requirements 
| Parameter | 
Description | 
| supported Windows operating systems | 
Windows 10, 11 on x86-64 | 
| supported macOS operating systems | 
macOS 10.11+ on x86-64 and ARMv8 | 
| supported Linux distributions | 
GNU/Linux (Ubuntu 14.04+, CentOS 7+, Debian 8+) on x86-64 and ARMv8 | 
| supported processors | 
x86-64 (Intel, AMD), ARMv8 (e.g., Raspberry Pi 4 and newer, Apple M-series) | 
Analog Interface Specifications
Table 4: Pump Tone Generation (Parametric Pump Channels)
| Parameter | 
Details | 
  | 
Min | 
Typ | 
Max | 
| Output impedance | 
- | 
  | 
- | 
50 Ω | 
- | 
| Output coupling | 
- | 
  | 
AC | 
  | 
  | 
| Synthesizer frequency range | 
Pump Tone | 
  | 
- | 
4...12 GHz | 
2.5...12 GHz | 
|   | 
Probe Tone | 
  | 
- | 
4...8.5 GHz | 
2.5...10 GHz | 
| Synthesizer frequency step | 
- | 
  | 
1 kHz | 
- | 
- | 
| Nominal output level range | 
Pump Tone | 
HIGH | 
- | 
0...20 dBm | 
- | 
|   | 
  | 
LOW | 
- | 
-20...0 dBm | 
- | 
|   | 
Probe Tone | 
at 6 GHz | 
- | 
-35...-50 dBc | 
- | 
| Maximum settable output level | 
Pump Tone, ALC enabled and low-pass filter disabled | 
4…8 GHz | 
- | 
20 dBm | 
- | 
|   | 
  | 
8…9 GHz | 
- | 
18 dBm | 
- | 
|   | 
  | 
9…10 GHz | 
- | 
14 dBm | 
- | 
|   | 
  | 
10…11 GHz | 
- | 
12 dBm | 
- | 
|   | 
  | 
11…12 GHz | 
- | 
7 dBm | 
- | 
| Output level error | 
Pump Tone | 
  | 
- | 
1.5 dB | 
- | 
|   | 
Probe Tone | 
  | 
- | 
3 dB | 
- | 
| Output level step | 
Pump Tone | 
ALC enabled | 
- | 
0.01 dB | 
- | 
|   | 
Probe Tone | 
  | 
- | 
0.5 dB | 
- | 
| Maximum output level drift with temperature | 
Pump Tone | 
ALC enabled | 
- | 
0.02 dB / °C | 
- | 
| Maximum output level drift over time | 
Pump Tone | 
ALC enabled, at 25 °C | 
- | 
0.02 dB / 24h | 
- | 
| Output phase noise | 
10 GHz | 
offset 1 kHz | 
- | 
-95 dBc/Hz | 
- | 
|   | 
  | 
offset 10 kHz | 
- | 
-105 dBc/Hz | 
- | 
|   | 
  | 
offset 1 MHz | 
- | 
-115 dBc/Hz | 
- | 
|   | 
  | 
offset 10 MHz | 
- | 
-145 dBc/Hz | 
- | 
Table 5: Readout Signal Conditioning (Parametric Pump Channels)
| Parameter | 
Details | 
Min | 
Typ | 
Max | 
| Signal In impedance | 
- | 
- | 
50 Ω | 
- | 
| Signal In coupling | 
- | 
AC | 
  | 
  | 
| Signal Out impedance | 
- | 
- | 
50 Ω | 
- | 
| Signal Out coupling | 
- | 
AC | 
  | 
  | 
| Pump In impedance | 
- | 
- | 
50 Ω | 
- | 
| Pump In coupling | 
- | 
AC | 
  | 
  | 
| Signal In to Signal Out frequency range | 
- | 
- | 
5...9 GHz | 
- | 
| Signal In to Signal Out gain | 
- | 
- | 
10 dB | 
- | 
| Signal In to Signal Out noise figure | 
8 GHz | 
- | 
5 dB | 
- | 
| Signal In maximum level | 
8 GHz | 
- | 
-7 dBm | 
- | 
| Pump tone cancellation frequency range | 
- | 
- | 
5...10 GHz | 
- | 
| Cancellation tone attenuation span | 
- | 
- | 
50 dB | 
- | 
| Cancellation tone phase shift span | 
- | 
360 deg | 
- | 
- | 
Table 6: Other Inputs and Outputs
| Parameter | 
Details | 
Min | 
Typ | 
Max | 
| Reference clock input | 
- | 
SMA on back panel | 
  | 
  | 
| Reference clock input impedance | 
- | 
- | 
50 Ω | 
- | 
| Reference clock input coupling | 
- | 
AC | 
  | 
  | 
| Reference clock input frequency | 
- | 
10 / 100 MHz | 
  | 
  | 
| Reference clock input level range | 
- | 
- | 
-10...+14 dBm | 
- | 
| Reference clock output | 
- | 
SMA on back panel | 
  | 
  | 
| Reference clock output impedance | 
- | 
- | 
50 Ω | 
- | 
| Reference clock output coupling | 
- | 
AC | 
  | 
  | 
| Reference clock output frequency | 
- | 
10/100 MHz | 
  | 
  | 
| Reference clock output level | 
10MHz | 
- | 
+8 dBm | 
- | 
|   | 
100 MHz | 
- | 
+7 dBm | 
- |